FDA Adverse Event Injury Summary report: N

SROM STM STD 36+12L 17X22

MDR report key: 2834302 · Received November 15, 2012

Report

Report Number
1818910-2012-27433
Event Type
Injury
Date Received
November 15, 2012
Date of Event
September 27, 2011
Report Date
October 30, 2012
Manufacturer
DEPUY RAYNHAM
Product Code
MRA
PMA / PMN Number
P040023
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EXAMINATION OF THE REPORTED DEVICE WAS NOT POSSIBLE AS IT WAS NOT RETURNED. A SEARCH OF THE COMPLAINTS DATABASES FINDS NO OTHER REPORTS AGAINST THE PRODUCT AND LOT CODE COMBINATION SINCE ITS RELEASE TO DISTRIBUTION. THE INVESTIGATION CAN DRAW NO CONCLUSION WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION HAS NOT BEEN INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Description of Event or Problem · 1

PATIENT WAS REVISED DUE TO PAIN, FLUID AND METALLOSIS. IT WAS NOTED THAT THE COBALT LEVEL WAS 3.6 AND THE CHROMIUM LEVEL WAS 1.9. LITIGATION PAPERS RECEIVED (B)(4) 2012 AND ATTACHED. COMPLAINT CHANGED TO LEGAL. THERE IS NO NEW INFORMATION THAT WOULD CHANGE THE OUTCOME OF THE INVESTIGATION. (B)(4) 2012 - PLAINTIFF'S PRELIMINARY DISCLOSURE FORM WAS RECEIVED, WHICH IDENTIFIED (PART/LOT) INFORMATION. THE COMPLAINT AND ASSOCIATED MDRS WERE UPDATED. THERE WAS NO NEW INFORMATION THAT WOULD CHANGE THE OUTCOME OF THE INVESTIGATION. (B)(4) 2012 - PATIENT'S OPERATIVE RECORDS RECEIVED. NOTED WERE FINE LINEAR SCRATCHES, WHICH WERE MULTIDIRECTIONAL, ON THE FEMORAL HEAD. THE TAPER HAD CORROSIVE DEBRIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SROM STM STD 36+12L 17X22 FEMORAL STEM MRA DEPUY RAYNHAM 2844684

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention