FDA Adverse Event Malfunction Summary report: N

SEE H-10

MDR report key: 2834279 · Received November 15, 2012

Report

Report Number
2023826-2012-00926
Event Type
Malfunction
Date Received
November 15, 2012
Date of Event
August 19, 2010
Report Date
November 14, 2012
Manufacturer
STAAR SURGICAL COMPANY
Product Code
HQL
PMA / PMN Number
P880091
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULAR LENS (ELASTIMIDE). (B)(4) NO LENS IN RETURNED CARTON. CONCLUSION: AN INVESTIGATION OF THE ROOT CAUSES OF LENS TEARS, SIMILAR TO THAT STATED IN THIS CLAIM, WERE ADDRESSED IN A CAPA OPENED, WHICH WAS SUBSEQUENTLY CLOSED. THE CAPA ADDRESSED DELIVERY SYSTEM ISSUES INCLUDING ALL STAGES OF MANUFACTURING OF THE INJECTORS AND CARTRIDGES. PROCESSES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. THE CAPA ALSO ADDRESSED HANDLING ERRORS. ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. BASED ON THE COMPLAINT HISTORY AND THE EVALUATION OF THE RETURNED PRODUCT, POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND HANDLING ERRORS BY THE CUSTOMER. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HAPTIC OF AN AQ2010V SILICONE THREE PIECE LENS TORE OFF DURING INSERTION. THERE WAS NO PATIENT INJURY. NO FURTHER INFORMATION WAS AVAILABLE ON THIS INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SEE H-10 INTRAOCULAR LENS HQL STAAR SURGICAL COMPANY AQ2010V N/A

Patients

Seq Age Sex Outcome Treatment
1 8 YR