INFUSOR
Report
- Report Number
- 1416980-2012-05300
- Event Type
- Malfunction
- Date Received
- November 15, 2012
- Date of Event
- October 22, 2012
- Report Date
- October 22, 2012
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). BAXTER RECEIVED ONE SAMPLE FOR EVALUATION. THE REPORTED CONDITION OF A LEAK WAS NOT CONFIRMED. VISUAL EXAMINATION OF THE SAMPLE SHOWED NO SIGNS OF PHYSICAL ABNORMALITY. A FUNCTIONAL LEAK TEST WAS PERFORMED ON THE PCM. AT THE NORMAL OPERATION SPECIFICATION OF 25 PSI, NO SIGNS OF LEAKAGE WERE OBSERVED ON ANY PART OF THE DEVICE. NO ROOT CAUSE WAS IDENTIFIED, AS NO EVIDENCE OF PRODUCT MALFUNCTION WAS OBSERVED. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT THIS PRODUCT IS NOT DISTRIBUTED IN THE US AND DOES NOT HAVE A 510(K) NUMBER. THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION.
BAXTER (B)(4) RECEIVED A REPORT THAT A PATIENT CONTROL MODULE (PCM) LEAKED FROM THE RESERVOIR DURING FILLING. THIS CONDITION HAS THE POTENTIAL TO INTERRUPT THERAPY OR BREACH THE STERILE FLUID PATHWAY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFUSOR | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - IRVINE | 12C010 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |