PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2012-07200
- Event Type
- Injury
- Date Received
- November 15, 2012
- Date of Event
- October 12, 2012
- Report Date
- October 30, 2012
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED EVENT WAS CONFIRMED. BASED ON VISUAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION. THE SECOND PERCLOSE PROGLIDE DEVICE IS BEING FILED UNDER A SEPARATE MEDWATCH MFR NUMBER. IT WAS REPORTED THAT THE PROGLIDE DEVICE WAS USED IN A CALCIFIED VESSEL. PER THE INSTRUCTIONS FOR USE, THE SAFETY AND EFFECTIVENESS OF THE PROGLIDE DEVICE HAVE NOT BEEN ESTABLISHED FOR PATIENTS WITH FEMORAL ARTERY CALCIUM WHICH IS FLUOROSCOPICALLY VISIBLE AT THE ACCESS SITE.
IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF A MODERATELY CALCIFIED COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PERCLOSE PROGLIDE DEVICE AFTER AN INTERVENTIONAL ILIAC STENTING PROCEDURE. REPORTEDLY, A CUFF MISS OCCURRED. A SECOND PERCLOSE PROGLIDE DEVICE WAS USED AND THERE WAS NO BLOOD RETURN. HEMOSTASIS WAS ACHIEVED BY APPLYING MANUAL ARTERIAL COMPRESSION. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA OR CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. THE PHYSICIAN IS REPORTEDLY TRAINED IN THE USE OF THE PERCLOSE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 20910J1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Required Intervention | SHEATH: 6F |