FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM ENDO CLIP APPLIER

MDR report key: 2834067 · Received November 15, 2012

Report

Report Number
3005075853-2012-05183
Event Type
Malfunction
Date Received
November 15, 2012
Date of Event
November 7, 2012
Report Date
November 8, 2012
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION IS UNAVAILABLE; DEVICE WAS NOT RETURNED FOR EVALUATION. DEVICE NOT RETURNED THE COMPLAINT COULD NOT BE CONFIRMED BECAUSE NO DEVICE WAS RETURNED FOR ANALYSIS. THE DEVICE HISTORY RECORDS WERE REVIEWED AND THE MANUFACTURING CRITERIA WAS MET PRIOR TO THE RELEASE OF THE BATCHES INCLUDED IN THIS LOT.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE FIRST CLIP WAS NOTED TO BE MISFEED INTO THE JAW OF THE DEVICE, AND UPON FORMING IT, IT SCISSORED. THE REMAINING CLIPS FED, AND FORMED AS INTENDED. IN ORDER TO CONFIRM THE CLIPS WERE WITHIN MANUFACTURING SPECIFICATIONS, THE CLIPS WERE EVALUATED USING A TOOL THAT IS DESIGNED TO DETERMINE PROPER FORMATION. DURING ANALYSIS THE JAWS OPEN AND CLOSE AS INTENDED. IN ADDITION, THE DEVICE LOCKED OUT AS INTENDED. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A CHOLECYSTECTOMY PROCEDURE, THE CLIPS ARE DELIVERED AS CROSSED. THE CLIPS WOULD NOT CLOSE AND WERE DELIVERED AS CROSSED. THIS PROBLEM OCCURRED AT THE FIRST CLIP AND THREE TIMES. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGAMAX-5MM ENDO CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA J4C672

Patients

Seq Age Sex Outcome Treatment
1