ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2012-13662
- Event Type
- Malfunction
- Date Received
- November 15, 2012
- Report Date
- October 23, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(4): TESTING CONFIRMED THE UP, OK, AND DOWN BUTTONS ARE INTERMITTENTLY RESPONSIVE. THE KEYPAD WAS INTACT AND WHEN REMOVED, CONTAMINATION WAS FOUND UNDER ALL CONTACTS. THE PUMP FAILED A LEAK TEST WITH BATTERY COMPARTMENT LEAK AND CRACK. MOISTURE WAS SEEN IN THE BATTERY COMPARTMENT.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2012, THE DISTRIBUTOR CONTACTED ANIMAS ALLEGING THAT THE PUMP HAS UNRESPONSIVE UP/DOWN KEYPAD BUTTONS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |