ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2012-13655
- Event Type
- Malfunction
- Date Received
- November 15, 2012
- Report Date
- October 19, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RI, US
- Reporter Occupation
- OTHER
Narratives
DEVICE EVALUATION SUBMITTED (B)(4) 2013 FOLLOW-UP # 1 - THE PUMP WAS RETURNED TO ANIMAS AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012: NO DEFECT WAS FOUND. THE BLACK BOX REVIEW SHOWS A MANUAL TIME/DATE CORRECTION ON THE REPORTED DATE FROM (B)(6) 2012. NO ACTIVITY OUTSIDE NORMAL USE OBSERVED IN THE BLACK BOX OR HISTORY DOWNLOAD DAILY INSULIN DELIVERY TOTALS CORRECTLY REFLECTED THE USER'S PROGRAMMED BASAL RATES. THE PUMP BOOTS UP NORMALLY, PERFORMED MULTIPLE "EZ-PRIME" OPERATIONS SUCCESSFULLY AND EXERCISED THE PUMP FOR 6 DAYS. NO ALARMS DUPLICATED DURING TESTING .THE UNIT PASSES "TIME TEST" SUCCESSFULLY AND FOUND ABLE TO KEEP "TIME/DATE" SETTINGS ACCURATELY DURING TESTING. OPENED THE PUMP TO INVESTIGATE, NO DAMAGE OR MOISTURE INGRESS WERE FOUND TO THE POWER CIRCUITS OR THE PCB.
THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2012, THE REPORTER CONTACTED ANIMAS FOR AN UNRELATED ISSUE AND DURING TROUBLESHOOTING FOUND THAT THE PUMP'S TIME WAS CORRECT BUT THE DATE WAS INCORRECT IN THE PUMP HISTORIES. THE REPORTER NOTED THAT THE PUMP'S DATE FOR THAT DAY'S RECORDS WAS (B)(6) 2012, WHEN THE DATE WAS REALLY (B)(6) 2012. THE REPORTER COULD NOT EXPLAIN HOW THE DATE COULD HAVE BECOME OFF BY A DAY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 14 YR |