FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2833995 · Received November 15, 2012

Report

Report Number
2531779-2012-13655
Event Type
Malfunction
Date Received
November 15, 2012
Report Date
October 19, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUBMITTED (B)(4) 2013 FOLLOW-UP # 1 - THE PUMP WAS RETURNED TO ANIMAS AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012: NO DEFECT WAS FOUND. THE BLACK BOX REVIEW SHOWS A MANUAL TIME/DATE CORRECTION ON THE REPORTED DATE FROM (B)(6) 2012. NO ACTIVITY OUTSIDE NORMAL USE OBSERVED IN THE BLACK BOX OR HISTORY DOWNLOAD DAILY INSULIN DELIVERY TOTALS CORRECTLY REFLECTED THE USER'S PROGRAMMED BASAL RATES. THE PUMP BOOTS UP NORMALLY, PERFORMED MULTIPLE "EZ-PRIME" OPERATIONS SUCCESSFULLY AND EXERCISED THE PUMP FOR 6 DAYS. NO ALARMS DUPLICATED DURING TESTING .THE UNIT PASSES "TIME TEST" SUCCESSFULLY AND FOUND ABLE TO KEEP "TIME/DATE" SETTINGS ACCURATELY DURING TESTING. OPENED THE PUMP TO INVESTIGATE, NO DAMAGE OR MOISTURE INGRESS WERE FOUND TO THE POWER CIRCUITS OR THE PCB.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2012, THE REPORTER CONTACTED ANIMAS FOR AN UNRELATED ISSUE AND DURING TROUBLESHOOTING FOUND THAT THE PUMP'S TIME WAS CORRECT BUT THE DATE WAS INCORRECT IN THE PUMP HISTORIES. THE REPORTER NOTED THAT THE PUMP'S DATE FOR THAT DAY'S RECORDS WAS (B)(6) 2012, WHEN THE DATE WAS REALLY (B)(6) 2012. THE REPORTER COULD NOT EXPLAIN HOW THE DATE COULD HAVE BECOME OFF BY A DAY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 14 YR