FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 2833970
·
Received November 9, 2012
Report
- Report Number
- 1720753-2012-09213
- Event Type
- Malfunction
- Date Received
- November 9, 2012
- Date of Event
- October 19, 2012
- Report Date
- November 9, 2012
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE REPRESENTATIVE EVALUATED THE SYSTEM AND ADJUSTED THE 5 VOLT POWER SUPPLY AND REPAIRED A BROKEN WIRE. THE SYSTEM OPERATES AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE SYSTEM EXPERIENCED COMMUNICATION ERRORS. THIS ERROR IS LIKELY TO RESULT IN A SYSTEM LOCK UP OR PREVENT THE SYSTEM FROM BOOTING TO USABLE STATE. THERE IS NO REPORT OF PATIENT INJURY OR DEATH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |