FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2833970 · Received November 9, 2012

Report

Report Number
1720753-2012-09213
Event Type
Malfunction
Date Received
November 9, 2012
Date of Event
October 19, 2012
Report Date
November 9, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND ADJUSTED THE 5 VOLT POWER SUPPLY AND REPAIRED A BROKEN WIRE. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM EXPERIENCED COMMUNICATION ERRORS. THIS ERROR IS LIKELY TO RESULT IN A SYSTEM LOCK UP OR PREVENT THE SYSTEM FROM BOOTING TO USABLE STATE. THERE IS NO REPORT OF PATIENT INJURY OR DEATH.