PROMUS ELEMENT¿ PLUS
Report
- Report Number
- 2134265-2012-06945
- Event Type
- Injury
- Date Received
- November 15, 2012
- Date of Event
- October 17, 2012
- Report Date
- October 19, 2012
- Manufacturer
- BOSTON SCIENTIFIC - GALWAY
- Product Code
- NIQ
- PMA / PMN Number
- P110010
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE EVALUATED BY MANUFACTURER ¿ A VISUAL AND MICROSCOPIC EXAMINATION OF THE DEVICE FOUND THE DEVICE WAS RETURNED TO THE COMPLAINT INVESTIGATION SITE WITH THE STENT DETACHED FROM THE BALLOON. THE BALLOON WAS FOUND TO HAVE THE STENT IMPRESSIONS ON IT AND PILLOWING OF THE BALLOON INDICATING THAT THE STENT WAS CRIMPED PER PROCESS IN THE CORRECT LOCATION DURING MANUFACTURING. THE STENT WAS ALSO RETURNED SEVERELY STRETCHED AND DAMAGED. KINKS WERE ALSO NOTED ALONG THE SHAFT OF THE DEVICE. THE BALLOON AND TIP SECTIONS OF THE DEVICE WERE VISUALLY AND MICROSCOPICALLY EXAMINED AND NO ISSUES WERE NOTED WITH THEIR PROFILES THAT COULD HAVE POTENTIALLY CONTRIBUTED TO THE COMPLAINT INCIDENT. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO POSITIVE PRESSURE. SOLIDIFIED CONTRAST MEDIA WAS PRESENT WITHIN THE ENTIRE LENGTH OF THE INFLATION LUMEN, THEREFORE INDICATING THE DEVICE HAD BEEN PREPPED FOR USE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).
DEVICE IS A COMBINATION PRODUCT. (B)(4).
IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE, A STENT EXPLANT OCCURRED. THE LESION BEING TREATED WAS LOCATED IN THE RIGHT CORONARY ARTERY (RCA). THE PHYSICIAN IMPLANTED A 2.74X24MM PROMUS ELEMENT PLUS STENT. THE STENT DELIVERY SYSTEM WAS REMOVED. HOWEVER, WHEN THE PHYSICIAN REMOVED THE 2 NON-BSC GUIDE WIRES THE STENT WAS EXPLANTED AND REMOVED WITH THE GUIDE WIRES. TWO NON-BSC STENTS WERE IMPLANTED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS OK.
IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE, A STENT EXPLANT OCCURRED. THE LESION BEING TREATED WAS LOCATED IN THE RIGHT CORONARY ARTERY (RCA). THE PHYSICIAN IMPLANTED A 2.74X24MM PROMUS ELEMENT PLUS STENT. THE STENT DELIVERY SYSTEM WAS REMOVED. HOWEVER, WHEN THE PHYSICIAN REMOVED THE 2 NON-BSC GUIDE WIRES THE STENT WAS EXPLANTED AND REMOVED WITH THE GUIDE WIRES. TWO NON-BSC STENTS WERE IMPLANTED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS OK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PROMUS ELEMENT¿ PLUS | STENT, CORONARY, DRUG-ELUTING | NIQ | BOSTON SCIENTIFIC - GALWAY | H7493911424270 | 14860620 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | BMW GUIDE WIRE X2| HOCKEYSTICK GUIDE CATHETER |