FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 2833651 · Received November 12, 2012

Report

Report Number
1627487-2012-11666
Event Type
Injury
Date Received
November 12, 2012
Date of Event
October 17, 2012
Report Date
October 17, 2012
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. MFR'S EVAL: CORRECTIVE AND PREVENTION ACTION (CAPA) INVESTIGATION WAS PERFORMED. RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND / OR CHARGING WAND OF SUFFICIENT ELEVATED TEMP TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REFERENCE MFR REPORT: 1627487-2012-11665. THE PT REPORTED HE WAS EXPERIENCING HEATING WHILE CHARGING AFTER 20 MINS, AND WAS UNABLE TO CHARGE THE IPG UNTIL IT WAS FULLY CHARGED. IN ADDITION, THE PT WAS CONCERNED HIS IPG MAY BE SUPERFICIAL. ATTEMPTS TO CONTACT THE PT WERE UNSUCCESSFUL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CHARGING SYSTEM SCS CHARGING SYSTEM LGW ST. JUDE MEDICAL - NEUROMODULATION 3711 3052207

Patients

Seq Age Sex Outcome Treatment
1 50 YR Required Intervention SCS EXTENSION: MODEL 3342| IMPLANT:| IMPLANT:| IMPLANT:| SCS LEAD: MODEL 3146 (2)| SCS ANCHOR: MODEL 1192| SCS LEAD: MODEL 3186| IMPLANT: