FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2833604 · Received October 29, 2012

Report

Report Number
1720753-2012-08793
Event Type
Malfunction
Date Received
October 29, 2012
Date of Event
October 19, 2012
Report Date
October 29, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SVC REP PERFORMED AN ONSITE INVESTIGATION. THE COLLIMATOR WAS RECALIBRATED. THE SYS WAS THEN FOUND TO BE OPERATING AS INTENDED. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYS DISPLAYED A COLLIMATOR IRIS TOO LARGE ERROR MESSAGE DURING PT IMAGING CASES. THERE IS NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1