FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2833588 · Received October 29, 2012

Report

Report Number
1720753-2012-08783
Event Type
Malfunction
Date Received
October 29, 2012
Date of Event
October 3, 2012
Report Date
October 29, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE WORKSTATION POWER SUPPLIES WERE REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE FIELD ENGINEER REPORTED THAT THE WORKSTATION WOULD NOT BOOT UP. THERE IS NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1