DIMENSION VISTA® CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 2517506-2012-00303
- Event Type
- Malfunction
- Date Received
- November 15, 2012
- Date of Event
- October 16, 2012
- Report Date
- October 16, 2012
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
- Product Code
- CIC
- PMA / PMN Number
- K061792
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY DEPRESSED CALCIUM RESULT IS SAMPLE INTEGRITY. THE SIEMENS HEALTHCARE DIAGNOSTICS TECHNICAL SERVICE CENTER REPRESENTATIVE DIRECTED THE CUSTOMER TO PERFORM ROUTINE MAINTENANCE PROCEDURES TO ADDRESS THE ISSUE. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
A FALSELY DEPRESSED CALCIUM RESULT WAS OBTAINED ON QC AND A PATIENT SAMPLE. IT IS UNKNOWN IF THE RESULT WAS REPORTED TO THE PHYSICIAN. A REPEAT WAS RUN AND A HIGHER CALCIUM RESULT WAS OBTAINED. IT IS UNKNOWN IF PATIENT TREATMENT WAS ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY DEPRESSED PATIENT CALCIUM RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION VISTA® CLINICAL CHEMISTRY SYSTEM | CALCIUM FLEX® REAGENT CARTRIDGE | CIC | SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW | 12185AA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |