FDA Adverse Event Malfunction Summary report: N

6800

MDR report key: 2833380 · Received October 26, 2012

Report

Report Number
1720753-2012-08766
Event Type
Malfunction
Date Received
October 26, 2012
Date of Event
October 17, 2012
Report Date
October 26, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED THE BACKPLANE, THE 5 VOLT POWER SUPPLY, AND THE X-RAY CONTROLLER BOARD. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT BOOT UP TO FULL FUNCTIONALITY. THERE IS NO REPORT OF PT INJURY OR DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 6800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 6800

Patients

Seq Age Sex Outcome Treatment
1