FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2833270 · Received October 26, 2012

Report

Report Number
1720753-2012-08726
Event Type
Malfunction
Date Received
October 26, 2012
Date of Event
October 15, 2012
Report Date
October 26, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE IDENTIFIED OR DUPLICATED. THE SYSTEM WAS OPERATING AS INTENDED. HOWEVER, AS A PRECAUTIONARY MEASURE, THE SERVICE REP ADJUSTED THE PS2 POWER SUPPLY VOLTAGE AND RESEATED THE GIB AND FFB BOARDS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM DISPLAYED A WHITE SCREEN AND WOULD NOT DISPLAY A USABLE FLUOROSCOPIC LIVE IMAGE. THERE IS NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1