FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2833247
·
Received November 15, 2012
Report
- Report Number
- 2032227-2012-07727
- Event Type
- Malfunction
- Date Received
- November 15, 2012
- Date of Event
- September 13, 2012
- Report Date
- November 8, 2012
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP WAS UNABLE TO PRIME DURING PRIME TEST AND ALARMED DURING BASIC OCCLUSION TEST DUE TO LOOSE DRIVE SUPPORT DISK. UNIT HAD CRACKED RESERVOIR TUBE AND BATTERY TUBE THREADS, NOTED DURING VISUAL INSPECTION.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP WAS NOT READING THE CORRECT AMOUNT OF INSULIN UNITS LEFT AND THE INSULIN SQUIRTING DURING REWIND. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | OYC | OYC | MEDTRONIC MINIMED | MMT-722NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 39 YR |