FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2833247 · Received November 15, 2012

Report

Report Number
2032227-2012-07727
Event Type
Malfunction
Date Received
November 15, 2012
Date of Event
September 13, 2012
Report Date
November 8, 2012
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS UNABLE TO PRIME DURING PRIME TEST AND ALARMED DURING BASIC OCCLUSION TEST DUE TO LOOSE DRIVE SUPPORT DISK. UNIT HAD CRACKED RESERVOIR TUBE AND BATTERY TUBE THREADS, NOTED DURING VISUAL INSPECTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP WAS NOT READING THE CORRECT AMOUNT OF INSULIN UNITS LEFT AND THE INSULIN SQUIRTING DURING REWIND. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP OYC OYC MEDTRONIC MINIMED MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 39 YR