ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2012-13638
- Event Type
- Malfunction
- Date Received
- November 15, 2012
- Report Date
- November 5, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(4): ON INVESTIGATION, THE KEYPAD WAS FOUND TO BE FULLY INTACT WITHOUT PEELING OR VISIBLE DAMAGE. ON TESTING, THE CONTRAST AND UP ARROW BUTTONS HAD INTERMITTENT RESPONSE. THE REMAINDER OF THE BUTTONS HAD NORMAL RESPONSE. THE KEYPAD COVER WAS REMOVED FOR INVESTIGATION AND REVEALED CONTAMINATION UNDER THE CONTACTS OF ALL THE KEYPAD BUTTONS.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE REPORTER CONTACTED ANIMAS ON (B)(6) 2012 AND STATED THE UP AND DOWN ARROW KEYPAD BUTTONS HAD DELAYED RESPONSES AND REQUIRED MULTIPLE PRESSES TO ELICIT A RESPONSE. THE REPORTER DENIED DAMAGE TO THE KEYPAD. THE PATIENT REPORTEDLY WORE THE PUMP UNDER A GARMENT, AGAINST THE BODY. THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED DUE TO THE ALLEGED KEYPAD RESPONSE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |