FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2833239 · Received November 15, 2012

Report

Report Number
2531779-2012-13638
Event Type
Malfunction
Date Received
November 15, 2012
Report Date
November 5, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): ON INVESTIGATION, THE KEYPAD WAS FOUND TO BE FULLY INTACT WITHOUT PEELING OR VISIBLE DAMAGE. ON TESTING, THE CONTRAST AND UP ARROW BUTTONS HAD INTERMITTENT RESPONSE. THE REMAINDER OF THE BUTTONS HAD NORMAL RESPONSE. THE KEYPAD COVER WAS REMOVED FOR INVESTIGATION AND REVEALED CONTAMINATION UNDER THE CONTACTS OF ALL THE KEYPAD BUTTONS.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

THE REPORTER CONTACTED ANIMAS ON (B)(6) 2012 AND STATED THE UP AND DOWN ARROW KEYPAD BUTTONS HAD DELAYED RESPONSES AND REQUIRED MULTIPLE PRESSES TO ELICIT A RESPONSE. THE REPORTER DENIED DAMAGE TO THE KEYPAD. THE PATIENT REPORTEDLY WORE THE PUMP UNDER A GARMENT, AGAINST THE BODY. THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED DUE TO THE ALLEGED KEYPAD RESPONSE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1 15 YR