FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2833213 · Received November 15, 2012

Report

Report Number
2032227-2012-07723
Event Type
Malfunction
Date Received
November 15, 2012
Date of Event
November 8, 2012
Report Date
November 8, 2012
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS UNABLE TO PRIME DURING THE PRIME TEST DUE TO A LOOSE MOTOR SUPPORT DISK. NO ALARMS WERE NOTED. THE INSULIN PUMP WAS RECEIVED WITH A MISSING END CAP STICKER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED DURING THE MANUAL PRIME. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP OYC OYC MEDTRONIC MINIMED MMT-722LCAH

Patients

Seq Age Sex Outcome Treatment
1