FDA Adverse Event
Malfunction
Summary report: N
PARADIGM INSULIN INFUSION PUMP
MDR report key: 2833211
·
Received November 15, 2012
Report
- Report Number
- 2032227-2012-07725
- Event Type
- Malfunction
- Date Received
- November 15, 2012
- Date of Event
- April 12, 2012
- Report Date
- November 8, 2012
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- K040676
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP HAD UNRESPONSIVE BUTTON DUE TO CORRODED KEYPAD TRACES. UNIT WAS UNABLE TO PERFORM THE DISPLACEMENT TEST DUE TO ANOMALY. UNIT HAD LOOSE DRIVE SUPPORT DISK, SCRATCHED DISPLAY WINDOW AND CRACK CASE AT DISPLAY WINDOW CORNERS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED DURING MANUAL PRIME. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM INSULIN INFUSION PUMP | LZG | OYC | MEDTRONIC MINIMED | MMT-715LNAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 24 YR |