FDA Adverse Event Malfunction Summary report: N

COATED VICRYL PLUS ANTIBACTERIAL (POLYGLACTIN 910)

MDR report key: 2833058 · Received November 15, 2012

Report

Report Number
2210968-2012-07348
Event Type
Malfunction
Date Received
November 15, 2012
Report Date
October 22, 2012
Manufacturer
ETHICON, INC.
Product Code
GAM
PMA / PMN Number
K032420
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY EVALUATED. THE EVALUATION REVEALED INDENTS AND SCUFF MARKS AROUND THE BREAK AREA PRODUCED DURING HANDLING BY A SURGICAL NEEDLE HOLDERS OR SOME OTHER GRIPPING DEVICE. THE NEEDLE FRACTURED DUE TO TENSILE OVERLOAD GENERATED DURING SEVERE MECHANICAL DEFORMATION, WITH SIGNS OF DUCTILE FAILURE. THERE ARE NO SIGNS OF MANUFACTURING DEFECTS FOUND ON THE NEEDLE. IN ORDER TO AVOID THIS KIND OF DAMAGE THE PACKAGE INSERT CAUTIONS: TO AVOID DAMAGING NEEDLE POINTS AND SWAGE AREAS, GRASP THE NEEDLE IN AN AREA ONE-THIRD (1/3) TO ONE-HALF (1/2) OF THE DISTANCE FROM THE SWAGED END TO THE POINT.

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL LAPAROSCOPIC HYSTERECTOMY ON AN UNKNOWN DATE AND SUTURE WAS USED. WHILE PERFORMING THE SUBCUTICULAR CLOSURE THE NEEDLE BROKE. THE PROCEDURE WAS COMPLETED WITH A LIKE DEVICE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COATED VICRYL PLUS ANTIBACTERIAL (POLYGLACTIN 910) SUTURE, ABSORBABLE GAM ETHICON, INC. NA NI

Patients

Seq Age Sex Outcome Treatment
1