COATED VICRYL PLUS ANTIBACTERIAL (POLYGLACTIN 910)
Report
- Report Number
- 2210968-2012-07348
- Event Type
- Malfunction
- Date Received
- November 15, 2012
- Report Date
- October 22, 2012
- Manufacturer
- ETHICON, INC.
- Product Code
- GAM
- PMA / PMN Number
- K032420
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4): THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY EVALUATED. THE EVALUATION REVEALED INDENTS AND SCUFF MARKS AROUND THE BREAK AREA PRODUCED DURING HANDLING BY A SURGICAL NEEDLE HOLDERS OR SOME OTHER GRIPPING DEVICE. THE NEEDLE FRACTURED DUE TO TENSILE OVERLOAD GENERATED DURING SEVERE MECHANICAL DEFORMATION, WITH SIGNS OF DUCTILE FAILURE. THERE ARE NO SIGNS OF MANUFACTURING DEFECTS FOUND ON THE NEEDLE. IN ORDER TO AVOID THIS KIND OF DAMAGE THE PACKAGE INSERT CAUTIONS: TO AVOID DAMAGING NEEDLE POINTS AND SWAGE AREAS, GRASP THE NEEDLE IN AN AREA ONE-THIRD (1/3) TO ONE-HALF (1/2) OF THE DISTANCE FROM THE SWAGED END TO THE POINT.
(B)(4). CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.
IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL LAPAROSCOPIC HYSTERECTOMY ON AN UNKNOWN DATE AND SUTURE WAS USED. WHILE PERFORMING THE SUBCUTICULAR CLOSURE THE NEEDLE BROKE. THE PROCEDURE WAS COMPLETED WITH A LIKE DEVICE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COATED VICRYL PLUS ANTIBACTERIAL (POLYGLACTIN 910) | SUTURE, ABSORBABLE | GAM | ETHICON, INC. | NA | NI |