GYNECARE MORCELLEX* TISSUE MORCELLATOR
Report
- Report Number
- 2210968-2012-07291
- Event Type
- Malfunction
- Date Received
- November 15, 2012
- Date of Event
- October 11, 2012
- Report Date
- October 22, 2012
- Manufacturer
- ETHICON, INC
- Product Code
- HET
- PMA / PMN Number
- K100280
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- PHYSICIAN
Narratives
IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA.
(B)(4). BLADE SEIZED/STOPPED. CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED HAS BEEN RECEIVED, HOWEVER, THE PRODUCT EVALUATION IS NOT YET COMPLETE. ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.
(B)(4): THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE RELATIVE ROTATION BETWEEN THE DRIVE TUBE AND INNER/OUTER SHEATH WAS NOT FROZEN. HOWEVER, EVALUATOR NOTICE A CUT ALONG THE PERIPHERY OF AIR TUBE. DUE TO THIS CONDITION, THE DEVICE MIGHT HAVE STOPPED ROTATING DURING THE PROCEDURE. THE EXACT ROOT CAUSE OF THIS CONDITION COULD NOT BE DETERMINED FROM THE RESULTS OF EVALUATION.
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC SUPRACERVICAL HYSTERECTOMY ON (B)(6) 2012. DURING THE PROCEDURE, THE DEVICE SEIZED UP AND STOPPED WORKING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GYNECARE MORCELLEX* TISSUE MORCELLATOR | LAPAROSCOPE, GYNECOLOGIC | HET | ETHICON, INC | NA | MT217153 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |