FDA Adverse Event Malfunction Summary report: N

GYNECARE MORCELLEX* TISSUE MORCELLATOR

MDR report key: 2833045 · Received November 15, 2012

Report

Report Number
2210968-2012-07291
Event Type
Malfunction
Date Received
November 15, 2012
Date of Event
October 11, 2012
Report Date
October 22, 2012
Manufacturer
ETHICON, INC
Product Code
HET
PMA / PMN Number
K100280
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA.

Additional Manufacturer Narrative · 1

(B)(4). BLADE SEIZED/STOPPED. CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED HAS BEEN RECEIVED, HOWEVER, THE PRODUCT EVALUATION IS NOT YET COMPLETE. ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.

Additional Manufacturer Narrative · 1

(B)(4): THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE RELATIVE ROTATION BETWEEN THE DRIVE TUBE AND INNER/OUTER SHEATH WAS NOT FROZEN. HOWEVER, EVALUATOR NOTICE A CUT ALONG THE PERIPHERY OF AIR TUBE. DUE TO THIS CONDITION, THE DEVICE MIGHT HAVE STOPPED ROTATING DURING THE PROCEDURE. THE EXACT ROOT CAUSE OF THIS CONDITION COULD NOT BE DETERMINED FROM THE RESULTS OF EVALUATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC SUPRACERVICAL HYSTERECTOMY ON (B)(6) 2012. DURING THE PROCEDURE, THE DEVICE SEIZED UP AND STOPPED WORKING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GYNECARE MORCELLEX* TISSUE MORCELLATOR LAPAROSCOPE, GYNECOLOGIC HET ETHICON, INC NA MT217153

Patients

Seq Age Sex Outcome Treatment
1