FDA Adverse Event Malfunction Summary report: N

PK DISSECTING FORCEPS INSTRUMENT

MDR report key: 2832869 · Received November 14, 2012

Report

Report Number
2955842-2012-00938
Event Type
Malfunction
Date Received
November 14, 2012
Date of Event
October 19, 2012
Report Date
October 17, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K061260
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. CUSTOMER INITIALLY REPORTED BROKEN CABLE. HOWEVER FOR CLARIFICATION, THE PRODUCT PROBLEM WAS A FRAYED PITCH CABLE. BASED ON THIS, THE COMPLAINT WAS CONFIRMED. FRAYED PITCH CABLE WAS FOUND AT DISTAL CLEVIS HUB. FRAYED SEGMENT ARE VARIOUS IN LENGTH. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER A DA VINCI SI HYSTERECTOMY PROCEDURE, THE SURGICAL STAFF REPORTED SEEING A BROKEN CABLE ON THE PK DISSECTING FORCEPS INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PK DISSECTING FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420227-03 M11120904 193

Patients

Seq Age Sex Outcome Treatment
1 43 YR DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES