FDA Adverse Event
Malfunction
Summary report: N
PK DISSECTING FORCEPS INSTRUMENT
MDR report key: 2832869
·
Received November 14, 2012
Report
- Report Number
- 2955842-2012-00938
- Event Type
- Malfunction
- Date Received
- November 14, 2012
- Date of Event
- October 19, 2012
- Report Date
- October 17, 2012
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K061260
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT WAS RETURNED AND EVALUATED. CUSTOMER INITIALLY REPORTED BROKEN CABLE. HOWEVER FOR CLARIFICATION, THE PRODUCT PROBLEM WAS A FRAYED PITCH CABLE. BASED ON THIS, THE COMPLAINT WAS CONFIRMED. FRAYED PITCH CABLE WAS FOUND AT DISTAL CLEVIS HUB. FRAYED SEGMENT ARE VARIOUS IN LENGTH. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT AFTER A DA VINCI SI HYSTERECTOMY PROCEDURE, THE SURGICAL STAFF REPORTED SEEING A BROKEN CABLE ON THE PK DISSECTING FORCEPS INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PK DISSECTING FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420227-03 | M11120904 193 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR | DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES |