COULTER® HMX HEMATOLOGY ANALYZER WITH AUTOLOADER
Report
- Report Number
- 1061932-2012-02668
- Event Type
- Malfunction
- Date Received
- November 14, 2012
- Date of Event
- October 22, 2012
- Report Date
- October 22, 2012
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- GKZ
- PMA / PMN Number
- K010765
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
PV21, NORMAL STATE: OPENS PATH FROM NEEDLE BELLOWS TO WASTE, VENT WASTE CHAMBER (VC3). IN OPERATED STATE: OPENS DRAIN PATH FROM VC3 TO VC7. PV49, NORMAL STATE: PROVIDES OPEN PATH FOR VACUUM AND DILUENT TO PM8. PROVIDES OPEN VACUUM PATH FROM PROBE WIPER TO VC7 VIA VL35. IN OPERATED STATE: OPENS PATH FOR 30-PSI PRESSURE TO PM8. OPENS DILUENT PATH FOR NEEDLE BACKWASH. PV40 PROVIDES OPEN DRAIN PATH FROM DIFF WASTE CHAMBER (VC7) TO WASTE CHAMBER (VC1). (B)(4).
THE CUSTOMER REPORTED APPROXIMATELY 3 ML OF UNCONTAINED DILUENT/BLOOD LEAK ON THE COULTER HMX HEMATOLOGY ANALYZER WITH AUTOLOADER AT THE PROBE WASH. THE FLUIDS ASSOCIATED WITH THE LEAK ARE BLOOD, DILUENT AND CLENZ. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT CONSISTING OF GLOVES AND LABORATORY COAT WHEN THE LEAK WAS FOUND. THE CUSTOMER DID NOT COME IN CONTACT WITH THE FLUID AND DID NOT SEEK MEDICAL ATTENTION. NO EXPOSURE (SPRAYED OR SPLASHED) TO MUCOUS MEMBRANES OR OPEN WOUNDS WERE REPORTED. NO DEATH, INJURY OR CHANGES TO PATIENT TREATMENT OR RESULTS ARE ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT REVIEW THE MSDS; HOWEVER, THE FACILITY DOES HAVE A RISK MANAGEMENT / EXPOSURE CONTROL PLAN IS IN PLACE. ON THE DAY OF THE EVENT ((B)(6) 2012) A FIELD SERVICE ENGINEER (FSE) WAS ON SITE AND ADJUSTED THE RINSE BLOCK AND REPLACED ALL TUBING THROUGH PV21, 49 AND 40, WHICH RESOLVED THE LEAK. THE FAILURE MODE OF THE LEAK WAS ATTRIBUTED TO PARTS REPLACED BY THE FSE DURING INSTRUMENT SERVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COULTER® HMX HEMATOLOGY ANALYZER WITH AUTOLOADER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER, INC. | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |