FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2832589 · Received November 14, 2012

Report

Report Number
3004209178-2012-90771
Event Type
Malfunction
Date Received
November 14, 2012
Date of Event
November 3, 2012
Report Date
November 3, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH A DETACHED END CAP, CRACKED LCD DISPLAY WINDOW CORNERS, BROKEN RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS AND MISSING END CAP STICKER.

Description of Event or Problem · 1

THE CUSTOMER STATED THAT THE PISTON IN THE RESERVOIR COMPARTMENT PUSHED OUT FROM THE BACKSIDE OF THE INSULIN PUMP, AND ALSO THE END CAP STICKER FELL OFF. ADVISED THE CUSTOMER TO DISCONTINUE THE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP OYC OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAL

Patients

Seq Age Sex Outcome Treatment
1 62 YR