FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2832554
·
Received November 14, 2012
Report
- Report Number
- 3004209178-2012-90776
- Event Type
- Malfunction
- Date Received
- November 14, 2012
- Date of Event
- September 6, 2012
- Report Date
- November 14, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE BASIC OCCLUSION TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK. HOWEVER, THE DEVICE PASSED THE PRIME TEST AND THE NUMBERS WERE NOT RAMPIN ON THEIR OWN. THE MOTOR PASSED THE MOTOR TEST.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT INSULIN SQUIRTED OUT AND THE NUMBERS WERE NOT RAMPING UP. THEN THE CUSTOMER CALLED BACK AND STATED THAT THE ROUND PLASTIC PART AT THE BOTTOM OF THE INSULIN PUMP WAS PUSHED OUT. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | OYC | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR |