FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2832554 · Received November 14, 2012

Report

Report Number
3004209178-2012-90776
Event Type
Malfunction
Date Received
November 14, 2012
Date of Event
September 6, 2012
Report Date
November 14, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE BASIC OCCLUSION TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK. HOWEVER, THE DEVICE PASSED THE PRIME TEST AND THE NUMBERS WERE NOT RAMPIN ON THEIR OWN. THE MOTOR PASSED THE MOTOR TEST.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT INSULIN SQUIRTED OUT AND THE NUMBERS WERE NOT RAMPING UP. THEN THE CUSTOMER CALLED BACK AND STATED THAT THE ROUND PLASTIC PART AT THE BOTTOM OF THE INSULIN PUMP WAS PUSHED OUT. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP OYC OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 53 YR