FDA Adverse Event
Malfunction
Summary report: N
LARGE NEEDLE DRIVER INSTRUMENT
MDR report key: 2832512
·
Received November 14, 2012
Report
- Report Number
- 2955842-2012-00927
- Event Type
- Malfunction
- Date Received
- November 14, 2012
- Date of Event
- October 16, 2012
- Report Date
- October 16, 2012
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE GRIP CABLE AT THE DISTAL IDLERS IS BROKE. THE CABLE HAS FRAYED SECTIONS THAT VARY IN LENGTH THAT ARE STICKING OUT AT THE WRIST. THE PULLEY DOES NOT EXHIBIT ANY DAMAGE. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A DA VINCI SI HYSTERECTOMY PROCEDURE, THE CABLE ON THE LARGE NEEDLE DRIVER INSTRUMENT SNAPPED. THERE WERE NO MISSING OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LARGE NEEDLE DRIVER INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420296-02 | M10120509 803 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SURGICAL SYSTEM AND ACCESSORIES |