FENESTRATED BIPOLAR FORCEPS INSTRUMENT

MDR report key: 2832472 · Received November 14, 2012

Report

Report Number
2955842-2012-00918
Event Type
Malfunction
Date Received
November 14, 2012
Date of Event
October 5, 2012
Report Date
October 16, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING WAS UNABLE TO CONFIRM THE CUSTOMER REPORTED FAILURE MODE. HOWEVER, THE PITCH CABLE AT THE DISTAL CLEVIS HUB IS BROKEN. THE CABLE SEGMENT THAT CONTAINS THE CRIMP IS STILL INSTALLED IN CLEVIS. THE CLEVIS DOES NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. THE INSTRUMENT PASSED ELECTRICAL CONTINUITY TESTING. ENGINEERING EVALUATION ALS FOUND THAT THE DISTAL END OF THE MAIN TUBE HAS VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. ENGINEERING CONCLUDED THAT DAMAGE TO THE MAIN TUBE WAS LIKELY CAUSED BY INSTRUMENT COLLISIONS OR ROUGH HANDLING. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS, O HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. O DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING SET UP OF A DA VINCI SI SURGICAL PROCEDURE, THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT WOULD NOT OPEN AND CLOSE. THERE WERE NO MISSING OR FALLEN PIECES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FENESTRATED BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420205-05 M10120313 941

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SURG SYS, ACCESSORIES & ESU