FDA Adverse Event Malfunction Summary report: N

MEGASUTURECUT NEEDLE DRIVER

MDR report key: 2832210 · Received November 14, 2012

Report

Report Number
2955842-2012-00922
Event Type
Malfunction
Date Received
November 14, 2012
Date of Event
October 18, 2012
Report Date
October 18, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT FRAYED CABLE SECTIONS WERE STICKING OUT AT THE DISTAL IDLER PULLEY. THE GRIP CABLE WAS ALSO OBSERVED TO BE DERAILED. THE GRIPS WERE ABLE TO OPEN AND CLOSE, BUT THEIR MOVEMENT AND FUNCTIONALITY COULD NOT BE PRECISE. EVIDENCE NOT CONCLUSIVE, BUT CABLE DERAILMENT WAS LIKELY DUE TO CABLE LOSING CONTACT WITH THE PULLEY DURING WRIST ARTICULATION. NO OTHER DAMAGE FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DAVINCI SI SURGICAL PROCEDURE, FRAYED CABLES WERE IDENTIFIED DURING SET UP ON THE MEGA SUTURE CUT NEEDLE DRIVER INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MEGASUTURECUT NEEDLE DRIVER ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420309-01 M10120427 640

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS| ACCESSORIES