FDA Adverse Event
Malfunction
Summary report: N
PK DISSECTING FORCEPS
MDR report key: 2832065
·
Received November 14, 2012
Report
- Report Number
- 2955842-2012-00920
- Event Type
- Malfunction
- Date Received
- November 14, 2012
- Date of Event
- October 17, 2012
- Report Date
- October 17, 2012
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K061260
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE PITCH CABLE WAS FOUND TO BE BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. NO OTHER DAMAGE FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SI SURGICAL PROCEDURE, THE CUSTOMER NOTED THAT 'BROKEN WIRE WAS STICKING OUT' ON THE PK DISSECTING FORCEPS INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PK DISSECTING FORCEPS | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420227-03 | M10120501 755 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES |