ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2012-13593
- Event Type
- Malfunction
- Date Received
- November 14, 2012
- Report Date
- November 3, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP # 1 DATE OF SUBMISSION 03/14/2013- DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE LIFTING NEAR THE UP ARROW BUTTON. A DAMAGED KEYPAD WILL ALLOW CONTAMINATION TO PERMEATE THE BUTTON CONTACTS NEGATIVELY IMPACTING THE BUTTON FUNCTION. THE DAMAGED KEYPAD IS OBVIOUS AND DETECTABLE AND SHOULD ALERT THE USER TO DISCONTINUE USE OF THE PUMP. DURING TESTING, THE UP AND DOWN ARROW BUTTONS WERE INTERMITTENTLY UNRESPONSIVE AND REQUIRED EXCESSIVE FORCE TO ACTIVATE; ALL OTHER BUTTONS RESPONDED APPROPRIATELY. DURING INVESTIGATION, EVIDENCE OF CONTAMINATION WAS FOUND UNDER ALL KEY CONTACTS. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED A DISCOLORED/FADED DISPLAY SCREEN, WHICH HAS NO EFFECT ON INSULIN DELIVERY FUNCTION.
ON (B)(6) /2012, THE REPORTER CONTACTED ANIMAS AND STATED THAT THE UP ARROW, DOWN ARROW AND OK KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE, REQUIRED MULTIPLE PRESSES TO ELICIT A RESPONSE AND OCCASIONALLY BECAME HYPER-SENSITIVE. THE REPORTER DENIED ANY DAMAGE TO THE RUBBER KEYPAD OR EVIDENCE OF MOISTURE IN THE PUMP. THE PATIENT REPORTEDLY WORE THE PUMP ATTACHED TO A BELT AND DID NOT CLEAN THE PUMP. THERE WAS NO REPORTED ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED DUE TO AN ALLEGED KEYPAD ISSUE.