FDA Adverse Event Malfunction Summary report: N

8800

MDR report key: 2832040 · Received November 14, 2012

Report

Report Number
1720753-2012-09304
Event Type
Malfunction
Date Received
November 14, 2012
Date of Event
October 15, 2012
Report Date
November 14, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EZ
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE VIDEO CONTROLLER PCB WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM FAILED TO INITIALIZE AND ALLOW FOR FLUOROSCOPIC EXPOSURES TO BE CREATED. NO PATIENT OR USER INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 8800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 8800

Patients

Seq Age Sex Outcome Treatment
1