FDA Adverse Event Injury Summary report: N

FREESTYLE AORTIC ROOT BIOPROSTHESIS

MDR report key: 2831994 · Received November 14, 2012

Report

Report Number
2025587-2012-00178
Event Type
Injury
Date Received
November 14, 2012
Date of Event
October 15, 2012
Report Date
October 18, 2012
Manufacturer
HEART VALVES SANTA ANA
Product Code
LWR
PMA / PMN Number
P970031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT ANALYSIS: UPON RECEIPT AT MEDTRONIC'S QUALITY LABORATORY, THE SEWING RING WAS EVERTED. NOTE: PER THE IFU, SECTION 10.2 "BIOPROSTHESIS IMPLANTATION- TAKE CARE NOT TO EVERT (ROLL OUTWARD) THE INFLOW END OF THE BIOPROSTHESIS WHEN SUTURING THE VALVE TO THE PATIENT'S ANNULUS. EVERSION COULD DAMAGE THE VALVE TISSUE." THERE IS ONE VISIBLE AREA OF A POSSIBLE PSEUDOANEURYSM ADJACENT TO THE RIGHT CORONARY. THE EDGES WERE ROLLED AND SMOOTH SUGGESTING "ADVENTITIAL HEMORRHAGE." A TISSUE SECTION IN THE NON-CORONARY SINUS AREA WAS THIN AND STRETCHED BUT NOT RUPTURED SUGGESTING A POSSIBLE PSEUDOANEURYSM. NOTE: THE OTHER TWO AREAS IN QUESTION COULD NOT BE IDENTIFIED. ALL LEAFLETS WERE SLIGHTLY STIFF BUT FLEXIBLE EXCEPT WHERE HOST TISSUE EXTENDED ON THE INFLOW AND OUTFLOW. ALL LEAFLETS WERE INTACT. ALL COMMISSURES WERE INTACT. THIN REMNANTS OF PANNUS EXTENDED 1 TO 3 MM ALONG THE TUNICA OF ALL CUSPS WITH A LARGE REMNANT OBSERVED IN THE LEFT RIGHT INFERIOR COAPTIVE AREA. PANNUS LINED THE INTIMAL SURFACE OF THE LEFT CORONARY SINUS WALL AND ANASTOMOSIS. PINK TO MAROON COLORED THROMBOTIC HOST TISSUE LINED THE BELLY OF THE RIGHT AND NON-CORONARY CUSPS. AN UNKNOWN AMOUNT OF HOST TISSUE APPEARED TO HAVE BEEN REMOVED ON THE INFLOW. RADIOGRAPHY SHOWED NO EVIDENCE OF MINERALIZATION IN THE VALVE AND/OR HOST TISSUE. CONCLUSION: THE DEVICE HISTORY RECORD WAS REVIEWED AND SHOWED THAT THIS PRODUCT MET ALL MANUFACTURING SPECIFICATION FOR PRODUCT RELEASED FOR DISTRIBUTION. NO ISSUES WERE IDENTIFIED THAT WOULD HAVE IMPACTED THIS EVENT. REDUCED PERFORMANCE OF THE VALVE IS ATTRIBUTED TO HOST TISSUE OVERGROWTH. THESE FINDINGS ARE GENERALLY CONSIDERED A PATIENT RELATED CONDITION. ADDITIONALLY, ANALYSIS FINDINGS INDICATE THAT IMPROPER IMPLANT TECHNIQUE MAY HAVE CONTRIBUTED TO THE CLINICAL OBSERVATION. (B)(6). (B)(4).

Description of Event or Problem · 1

MEDTRONIC RECEIVED INFORMATION THAT 22 MONTHS FOLLOWING THE IMPLANT OF THIS 29 MM AORTIC ROOT BIOPROSTHETIC VALVE (AORTIC ROOT REPLACEMENT WITH CONCOMITANT ASCENDING ARCH REPAIR), CT SCAN NOTED A SMALL PSEUDOANEURYSM AT THE PROXIMAL ANASTOMOSIS SITE. ALSO NOTED WERE 2 PSEUDOANEURYSMS ARISING FROM THE ANTERIOR, POSTERIOR ASPECTS OF THE ASCENDING AORTIC GRAFT WHICH SHOWED AN INCREASE IN SIZE FROM THE LAST SCAN 4 MONTHS PRIOR. THE PATIENT REMAINED ASYMPTOMATIC. DUE TO THE INCREASING SIZE OF THE PSEUDOANEURYSMS, SURGICAL INTERVENTION WAS PERFORMED (AORTIC ROOT RECONSTRUCTION WITH REPLACEMENT OF THE ASCENDING AORTA AND HEMIARCH WITH HOMOGRAFT). THE VALVE WAS EXPLANTED WITH NO ADVERSE PATIENT EFFECTS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FREESTYLE AORTIC ROOT BIOPROSTHESIS HEART-VALVE, NON-ALLOGRAFT TISSUE LWR HEART VALVES SANTA ANA FR995-29

Patients

Seq Age Sex Outcome Treatment
1 00073 YR Required Intervention