FREESTYLE AORTIC ROOT BIOPROSTHESIS
Report
- Report Number
- 2025587-2012-00178
- Event Type
- Injury
- Date Received
- November 14, 2012
- Date of Event
- October 15, 2012
- Report Date
- October 18, 2012
- Manufacturer
- HEART VALVES SANTA ANA
- Product Code
- LWR
- PMA / PMN Number
- P970031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT ANALYSIS: UPON RECEIPT AT MEDTRONIC'S QUALITY LABORATORY, THE SEWING RING WAS EVERTED. NOTE: PER THE IFU, SECTION 10.2 "BIOPROSTHESIS IMPLANTATION- TAKE CARE NOT TO EVERT (ROLL OUTWARD) THE INFLOW END OF THE BIOPROSTHESIS WHEN SUTURING THE VALVE TO THE PATIENT'S ANNULUS. EVERSION COULD DAMAGE THE VALVE TISSUE." THERE IS ONE VISIBLE AREA OF A POSSIBLE PSEUDOANEURYSM ADJACENT TO THE RIGHT CORONARY. THE EDGES WERE ROLLED AND SMOOTH SUGGESTING "ADVENTITIAL HEMORRHAGE." A TISSUE SECTION IN THE NON-CORONARY SINUS AREA WAS THIN AND STRETCHED BUT NOT RUPTURED SUGGESTING A POSSIBLE PSEUDOANEURYSM. NOTE: THE OTHER TWO AREAS IN QUESTION COULD NOT BE IDENTIFIED. ALL LEAFLETS WERE SLIGHTLY STIFF BUT FLEXIBLE EXCEPT WHERE HOST TISSUE EXTENDED ON THE INFLOW AND OUTFLOW. ALL LEAFLETS WERE INTACT. ALL COMMISSURES WERE INTACT. THIN REMNANTS OF PANNUS EXTENDED 1 TO 3 MM ALONG THE TUNICA OF ALL CUSPS WITH A LARGE REMNANT OBSERVED IN THE LEFT RIGHT INFERIOR COAPTIVE AREA. PANNUS LINED THE INTIMAL SURFACE OF THE LEFT CORONARY SINUS WALL AND ANASTOMOSIS. PINK TO MAROON COLORED THROMBOTIC HOST TISSUE LINED THE BELLY OF THE RIGHT AND NON-CORONARY CUSPS. AN UNKNOWN AMOUNT OF HOST TISSUE APPEARED TO HAVE BEEN REMOVED ON THE INFLOW. RADIOGRAPHY SHOWED NO EVIDENCE OF MINERALIZATION IN THE VALVE AND/OR HOST TISSUE. CONCLUSION: THE DEVICE HISTORY RECORD WAS REVIEWED AND SHOWED THAT THIS PRODUCT MET ALL MANUFACTURING SPECIFICATION FOR PRODUCT RELEASED FOR DISTRIBUTION. NO ISSUES WERE IDENTIFIED THAT WOULD HAVE IMPACTED THIS EVENT. REDUCED PERFORMANCE OF THE VALVE IS ATTRIBUTED TO HOST TISSUE OVERGROWTH. THESE FINDINGS ARE GENERALLY CONSIDERED A PATIENT RELATED CONDITION. ADDITIONALLY, ANALYSIS FINDINGS INDICATE THAT IMPROPER IMPLANT TECHNIQUE MAY HAVE CONTRIBUTED TO THE CLINICAL OBSERVATION. (B)(6). (B)(4).
MEDTRONIC RECEIVED INFORMATION THAT 22 MONTHS FOLLOWING THE IMPLANT OF THIS 29 MM AORTIC ROOT BIOPROSTHETIC VALVE (AORTIC ROOT REPLACEMENT WITH CONCOMITANT ASCENDING ARCH REPAIR), CT SCAN NOTED A SMALL PSEUDOANEURYSM AT THE PROXIMAL ANASTOMOSIS SITE. ALSO NOTED WERE 2 PSEUDOANEURYSMS ARISING FROM THE ANTERIOR, POSTERIOR ASPECTS OF THE ASCENDING AORTIC GRAFT WHICH SHOWED AN INCREASE IN SIZE FROM THE LAST SCAN 4 MONTHS PRIOR. THE PATIENT REMAINED ASYMPTOMATIC. DUE TO THE INCREASING SIZE OF THE PSEUDOANEURYSMS, SURGICAL INTERVENTION WAS PERFORMED (AORTIC ROOT RECONSTRUCTION WITH REPLACEMENT OF THE ASCENDING AORTA AND HEMIARCH WITH HOMOGRAFT). THE VALVE WAS EXPLANTED WITH NO ADVERSE PATIENT EFFECTS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREESTYLE AORTIC ROOT BIOPROSTHESIS | HEART-VALVE, NON-ALLOGRAFT TISSUE | LWR | HEART VALVES SANTA ANA | FR995-29 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00073 YR | Required Intervention |