DUROM ACETABULAR COMPONENT
Report
- Report Number
- 9613350-2012-01061
- Event Type
- Injury
- Date Received
- November 9, 2012
- Date of Event
- November 17, 2011
- Report Date
- October 23, 2012
- Manufacturer
- ZIMMER GMBH
- Product Code
- KWA
- Removal / Correction Number
- Z-2415/2426-2008
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIAN
Narratives
THE MANUFACTURER DID NOT RECEIVE DEVICES OR X-RAYS FOR REVIEW. WHERE LOT NUMBERS WERE RECEIVED FOR THE DEVICES, THE DEVICE HISTORY RECORDS WERE REVIEWED AND FOUND TO BE CONFORMING. BASED ON AN EXTENSIVE INVESTIGATION OF EVENTS REPORTED FROM SEVERAL USER FACILITIES OUTSIDE THE U.S., ZIMMER IDENTIFIED THAT THE MOST PROBABLE CAUSE FOR THE OUTCOME OBSERVED WAS A LOOSE OR UNSTABLE CUP THAT RESULTED FROM USE OF SURGICAL TECHNIQUES NOT CONSISTENT WITH THE MANUFACTURER'S RECOMMENDATIONS. AS A CORRECTIVE ACTION, A RETRAINING PROGRAM FOR USERS OUTSIDE THE U.S. WAS INITIATED IN NOVEMBER 2009 AND REPORTED TO THE NATIONAL COMPETENT AUTHORITIES AS REQUIRED. THE DUROM CUP REPORTED IN THIS CASE IS NOT MARKETED IN THE U.S. A SIMILAR CUP, COMPATIBLE WITH THE METASUL LDH FEMORAL HEAD, IS CLEARED IN THE U.S. AND A CORRECTIVE ACTION FOR THIS PRODUCT WAS REPORTED TO THE FDA IN JULY 2008 AS CORRECTION Z-2415/2426-2008. SHOULD ADDITIONAL INFORMATION INCLUDING THE FINAL INVESTIGATION RESULT BE AVAILABLE, THAT CHANGES THIS ASSESSMENT, AN AMENDED MEDICAL DEVICE REPORT WILL BE FILED. THE NEED FOR FURTHER CORRECTIVE MEASURES IS NOT INDICATED AT THIS TIME AND ZIMMER (B)(4) CONSIDERS THIS CASE AS CLOSED. (B)(4).
IT IS REPORTED THAT THE PT RECEIVED AN ACETABULAR CUP ON (B)(6) 2009. THE PT COMPLAINED OF NIGHT PAIN, RESTRICTED RANGE OF MOTION, IMPAIRED FUNCTION AND WALKING TIME BECAME LIMITED. THE PT UNDERWENT REVISION SURGERY ON (B)(6) 2011.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DUROM ACETABULAR COMPONENT | DUROM ACETABULAR COMPONENT AND METASUL | KWA | ZIMMER GMBH | 2479180 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | Hospitalization| R |