FDA Adverse Event
Malfunction
Summary report: N
IMMAGE® 800 IMMUNOCHEMISTRY SYSTEM
MDR report key: 2831562
·
Received November 14, 2012
Report
- Report Number
- 2050012-2012-01811
- Event Type
- Malfunction
- Date Received
- November 14, 2012
- Date of Event
- October 24, 2012
- Report Date
- October 24, 2012
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K962294
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE. THE FSE REPLACED FILTERS ON WASH STATIONS, CUVETTE WASH, AND HYDRO LINES 10 AND 16. THE FSE VERIFIED SYSTEM PERFORMANCE. FAILURE MODE DETERMINED TO BE PLUGGED WASTE LINE FILTERS. (B)(4).
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC., (BEC) AND REPORTED THAT SAMPLE AND REAGENT WASH STATIONS OVERFLOWED DUE TO PLUGGED WASTE LINE FILTERS ON IMMAGE 800 IMMUNOCHEMISTRY SYSTEM. NO ERROR MESSAGES, FLAGS OR QC OUTSIDE ACCEPTABLE LIMITS WERE OBSERVED PRIOR TO DISCOVERY OF THE LEAKING. THERE WAS NO EXPOSURE OR INJURIES TO THE OPERATORS. NO ERRONEOUS PATIENT RESULTS WERE GENERATED OR REPORTED DUE TO THIS ISSUE.