FDA Adverse Event Malfunction Summary report: N

DIMENSION® CLINICAL CHEMISTRY SYSTEM

MDR report key: 2831561 · Received November 14, 2012

Report

Report Number
1226181-2012-00104
Event Type
Malfunction
Date Received
November 14, 2012
Date of Event
October 15, 2012
Report Date
October 15, 2012
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC- BROOKFIELD
Product Code
JGS
PMA / PMN Number
K944093
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED SODIUM RESULT WAS A MISALIGNMENT OF THE INTEGRATED MULTISENSOR (IMT) PROBE. THE SIEMENS HEALTHCARE DIAGNOSTICS TECHNICAL SERVICE CENTER REPRESENTATIVE DIRECTED THE CUSTOMER TO TROUBLESHOOTING ACTIVITIES. THE CUSTOMER DISCOVERED THAT THE ALIGNMENT OF THE SAMPLE PROBE TO THE IMT PORT WAS OFF. THE CUSTOMER REPLACED THE SAMPLE PROBE WHICH RESOLVED THE PROBLEM. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

A FALSELY ELEVATED SODIUM RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE PATIENT RESULT WAS NOT REPORTED TO THE PHYSICIAN. THE RESULT WAS QUESTIONED WITHIN THE LABORATORY AND THE SAMPLE RE-RUN. THE SAMPLE WAS RE-RUN ON AN ALTERNATE INSTRUMENT AND A LOWER RESULT WAS OBTAINED. A CORRECTED RESULT WAS REPORTED.PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED SODIUM RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DIMENSION® CLINICAL CHEMISTRY SYSTEM CLINICAL CHEMISTRY SYSTEM JGS SIEMENS HEALTHCARE DIAGNOSTICS INC- BROOKFIELD EXL

Patients

Seq Age Sex Outcome Treatment
1