DIMENSION® CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 1226181-2012-00104
- Event Type
- Malfunction
- Date Received
- November 14, 2012
- Date of Event
- October 15, 2012
- Report Date
- October 15, 2012
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC- BROOKFIELD
- Product Code
- JGS
- PMA / PMN Number
- K944093
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED SODIUM RESULT WAS A MISALIGNMENT OF THE INTEGRATED MULTISENSOR (IMT) PROBE. THE SIEMENS HEALTHCARE DIAGNOSTICS TECHNICAL SERVICE CENTER REPRESENTATIVE DIRECTED THE CUSTOMER TO TROUBLESHOOTING ACTIVITIES. THE CUSTOMER DISCOVERED THAT THE ALIGNMENT OF THE SAMPLE PROBE TO THE IMT PORT WAS OFF. THE CUSTOMER REPLACED THE SAMPLE PROBE WHICH RESOLVED THE PROBLEM. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
A FALSELY ELEVATED SODIUM RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE PATIENT RESULT WAS NOT REPORTED TO THE PHYSICIAN. THE RESULT WAS QUESTIONED WITHIN THE LABORATORY AND THE SAMPLE RE-RUN. THE SAMPLE WAS RE-RUN ON AN ALTERNATE INSTRUMENT AND A LOWER RESULT WAS OBTAINED. A CORRECTED RESULT WAS REPORTED.PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED SODIUM RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION® CLINICAL CHEMISTRY SYSTEM | CLINICAL CHEMISTRY SYSTEM | JGS | SIEMENS HEALTHCARE DIAGNOSTICS INC- BROOKFIELD | EXL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |