FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2831464 · Received November 14, 2012

Report

Report Number
2531779-2012-13575
Event Type
Malfunction
Date Received
November 14, 2012
Report Date
November 7, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION (B)(4) 2012 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE FULLY INTACT; NO PEELING OR DAMAGE WAS OBSERVED. DURING TESTING, THE UP AND DOWN ARROW KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE; ALL OTHER BUTTONS RESPONDED APPROPRIATELY. DURING INVESTIGATION, EVIDENCE OF CONTAMINATION WAS FOUND UNDER ALL KEY CONTACTS. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED A CRACKED BATTERY COMPARTMENT. THE USER GUIDE WARNS THAT CRACKS, CHIPS, OR DAMAGE TO THE PUMP MAY IMPACT THE BATTERY CONTACT AND/OR THE WATERPROOF FEATURE OF THE PUMP.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

THE REPORTER CONTACTED ANIMAS ON (B)(6) 2012, STATING THAT THE OK KEYPAD BUTTON REQUIRED MULTIPLE PRESSES TO ELICIT A RESPONSE. THE REPORTER DENIED ANY DAMAGE TO THE KEYPAD OR ANY EVIDENCE OF MOISTURE IN THE PUMP. THE PATIENT REPORTEDLY WORE THE PUMP IN A POCKET AND CLEANED THE PUMP WITH A PAPER TOWEL. THERE IS NO ADVERSE EVENT WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED DUE TO THE ALLEGED KEYPAD ISSUES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 16 YR