FDA Adverse Event Malfunction Summary report: N

REDUCTION FORCEPS W/SERRATED JAW-LARGE HANDLE-S

MDR report key: 2831451 · Received October 11, 2012

Report

Report Number
8030965-2012-01022
Event Type
Malfunction
Date Received
October 11, 2012
Date of Event
September 13, 2012
Report Date
September 13, 2012
Manufacturer
SYNTHES GMBH
Product Code
HTD
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE USED FOR TREATMENT. MANUFACTURING EVALUATION RESULTS REVEALED THE DEVICE WAS RECEIVED WITH THE TIP BROKEN OFF. THE RELEVANT DIMENSIONS WERE MEASURED AND WERE FOUND WITHIN THE SPECIFICATION. THE FRACTURE FACE IS HOMOGENOUS WHICH INDICATES MATERIAL CONFORMITY. DUE TO THE EVENT DESCRIPTION AND THE EVALUATION OF THE RETURNED DEVICE, THE ROOT CAUSE IS UNABLE TO BE DETERMINED. THE PRODUCT DEVELOPMENT RECORDS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND. LOADING OF A HAND-HELD INSTRUMENT IS DICTATED BY FORCE EXERTED BY THE OPERATOR, NOT A FUNCTION OF DESIGN. EXCEEDING LOAD CAPACITY OF AN INSTRUMENT IS THEREFORE MISUSE. IMPOSSIBLE TO DETERMINE IF CAUSE OF FAILURE IS FROM INTENTIONAL MISUSE, PRODUCT DAMAGE OR EXCEEDING REASONABLE PRODUCT LIFE EXPECTANCY. PLACEHOLDER.

Additional Manufacturer Narrative · 1

THE MFG RECORDS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND. LOADING OF A HAND-HELD INSTRUMENT IS DICTATED BY FORCE EXERTED BY THE OPERATOR, NOT A FUNCTION OF DESIGN. EXCEEDING LOAD CAPACITY OF AN INSTRUMENT IS THEREFORE MISUSE. IMPOSSIBLE TO DETERMINE IF CAUSE OF FAILURE IS FROM INTENTIONAL MISUSE, PRODUCT DAMAGE OR EXCEEDING REASONABLE PRODUCT LIFE EXPECTANCY.

Description of Event or Problem · 1

DURING A PROCEDURE, THE END OF THE FORCEP BROKE OFF. ALL PARTS WERE RETRIEVED, ANOTHER SET OF FORCEPS WERE USED TO COMPLETE THE PROCEDURE. THERE WERE NO FURTHER INCIDENTS DURING THE PROCEDURE, NO DELAY TO THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 REDUCTION FORCEPS W/SERRATED JAW-LARGE HANDLE-S REDUCTION FORCEPS HTD SYNTHES GMBH 5005180

Patients

Seq Age Sex Outcome Treatment
1