LCP PROXTIBPI 3.5 LAT LE SHAFT 4HO L81 T
Report
- Report Number
- 3003506883-2012-00310
- Event Type
- Malfunction
- Date Received
- October 11, 2012
- Report Date
- September 12, 2012
- Manufacturer
- SYNTHES ELMIRA
- Product Code
- HRS
- PMA / PMN Number
- K030597
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
THIS DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS.ADDITIONAL NARRATIVE:THE AS RECEIVED CONDITION OF THE PLATE IS THAT THE SURFACE ANODIZE IS SLIGHTLY DISCOLORED FROM GOLD TO PINK CONSISTENT WITH NORMAL HANDLING. THERE ARE SOME SCRATCHES TO THE ANODIZED SURFACE IN THE NECK AREA OF THE PLATE AND ON THE TOP AND BOTTOM SURFACE OF THE HEAD. ALL THREADED AND NON-THREADED HOLES HAVE THE NORMAL ANODIZED SURFACE FINISH WITH NO DAMAGE EVIDENT. (B)(4): PLACEHOLDER.
THE RAW MATERIAL AND DEVICE HISTORY RECORDS WERE REVIEWED AND NOTHING WAS FOUND THAT WOULD CAUSE OR CONTRIBUTE TO THE REPORTED CONDITION.
THIS DEVICE IS USED FOR TREATMENT NOT DIAGNOSIS. THE INVESTIGATION FOUND THAT THE THREAD HOLES OF THE LCP PLATE CAN CUT THE USPU THREADS.
A DEVICE REPORT FROM (B)(6) INDICATED A HOSPITAL REPORTED: A RESPONSIBLE PERSON OF THE HOSPITAL FOUND THAT THREADS ON THE PLATE HOLES TEAR OFF/STRIPPED WHEN TIGHTENING. THERE WAS NO PATIENT INVOLVEMENT. THIS IS 5 OF 13 REPORTS FOR THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LCP PROXTIBPI 3.5 LAT LE SHAFT 4HO L81 T | LCP PROXTIBPLATE | HRS | SYNTHES ELMIRA | 6579705 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |