FDA Adverse Event Injury Summary report: N

ACHIEVA 1.5T NOVA

MDR report key: 2831132 · Received November 6, 2012

Report

Report Number
3003768277-2012-00081
Event Type
Injury
Date Received
November 6, 2012
Date of Event
October 21, 2012
Report Date
October 21, 2012
Manufacturer
PHILIPS HEALTHCARE
Product Code
LNH
PMA / PMN Number
K052013
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). (CONCLUSIONS) - THE INCIDENT REPORTED WAS CAUSED BY THE CLOTHING WORN BY THE PT. THE PT WAS NOT CHANGED IN HOSPITAL CLOTHING AND THE STRESS WORN BY THE PT CONTAINED EMBROIDERY WITH CONDUCTING MATERIAL. THE BURNS OCCURRED ON THE ABDOMEN PART CORRESPONDING TO THE LOCATION OF THE EMBROIDERY. THE INSTRUCTIONS FOR USE CONTAIN WARNING RELATED TO THIS ISSUE.

Description of Event or Problem · 1

THE CUSTOMER STATED A FEMALE PT WAS EXAMINED WITH A COMBINATION OF THE 15 CHANNEL SPINE AND 16 CHANNEL NEURO VASCULAR (NV) COIL. THERE WAS NOTHING ABNORMAL OBSERVED DURING THE EXAMINATION. THE PT WENT HOME AND INFORMED THE HOSPITAL AFTER OF A BURN, A SECOND DEGREE BURN OF 4X4 CAM WAS OBSERVED ON THE ABDOMEN AND CHEST.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACHIEVA 1.5T NOVA LNH PHILIPS HEALTHCARE 781196

Patients

Seq Age Sex Outcome Treatment
1 60 YR Other