FDA Adverse Event Malfunction Summary report: N

MATRIX THREADED PERSUADER- STANDARD

MDR report key: 2830979 · Received November 14, 2012

Report

Report Number
1719045-2012-01145
Event Type
Malfunction
Date Received
November 14, 2012
Date of Event
October 17, 2012
Report Date
October 17, 2012
Manufacturer
SYNTHES MONUMENT
Product Code
NKB
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. AS NOTED IN PRIOR COMPLAINTS, IMPROPER ASSEMBLY OF THE INSTRUMENT, NOT FULLY SEATING THE INSERT, BY THE OR STAFF LIKELY CAUSED THE REDUCTION INSERT TO SEE EXCESSIVE LOADING IN THE REGION OF FAILURE DURING REDUCTION. RATHER THAN HAVING THE AXIAL LOADS REACTED BY THE THRUST SURFACE ON THE UNDERSIDE OF THE HEAD, THE FINGERS INSTEAD CONTACT THE RETENTION RIB AND ARE SQUEEZED INTO THE INNER TUBE AS REDUCTION LOADS ARE APPLIED POTENTIALLY CAUSING THEM TO BREAK. THE BREAKAGE NOTED ON THE RETURNED PART IS CONSISTENT WITH THAT EXPECTED IF THE DEVICE IS ASSEMBLED INCORRECTLY AS NOTED. THE DESIGN RISK ASSESSMENT WAS REVIEWED AND WAS FOUND TO BE ADEQUATE FOR THE INTENDED USE.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. THE MANUFACTURING RECORDS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND. THE PRODUCT CONFORMED TO ALL REQUIREMENTS. PLACEHOLDER.

Description of Event or Problem · 1

DURING A L4-L5 PLIF PROCEDURE, THE THREADED INTERNAL BUSHING/PLASTIC THREADS OF THE MATRIX THREADED PERSUADER BROKE DURING THE REDUCTION. THE BROKEN PARTS WERE CONTAINED IN THE PERSUADER. THE SURGEON DID NOT USE AN ALTERNATE PERSUADER. THE SURGEON ACHIEVED REDUCTION WITH DISTRACTION. THERE WAS A TWENTY MINUTE DELAY IN THE PROCEDURE. THE PROCEDURE WAS COMPLETED. THERE ARE TWO PARTS, LEFT SIDE AND RIGHT SIDE, THEY COME AS A SINGLE SET YOU USE BOTH TOGETHER. BOTH DEVICES BROKE IN THIS PROCEDURE. THIS IS 2 OF 2 REPORTS FOR THE SAME EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MATRIX THREADED PERSUADER- STANDARD MATRIX THREADED PERSUADER NKB SYNTHES MONUMENT 6390270

Patients

Seq Age Sex Outcome Treatment
1 MATRIX THREADED PERSUADER