CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS

MDR report key: 2830834 · Received November 13, 2012

Report

Report Number
2015691-2012-18646
Event Type
Injury
Date Received
November 13, 2012
Date of Event
September 27, 2012
Report Date
October 16, 2012
Manufacturer
EDWARDS LIFESCIENCES
Product Code
DYE
PMA / PMN Number
P860057/S001
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE NOT RETURNED. UNFORTUNATELY, THE SOURCE OF THE REPORTED CALCIFICATION COULD NOT BE ASSESSED SINCE THE SUBJECT DEVICE WAS NOT EXPLANTED FROM THE PATIENT. ALTHOUGH THE ROOT CAUSE OF THIS EVENT CANNOT BE CONCLUSIVELY DETERMINED, IT APPEARS THAT THE STENOSIS WAS MOST LIKELY CAUSED BY THE CALCIFICATION IN THIS CASE. CALCIFICATION IS A WELL RECOGNIZED FAILURE MODE OF BIOPROSTHETIC VALVES. THE MECHANISMS FOR BIOPROSTHETIC HEART VALVE TISSUE CALCIFICATION ARE NOT FULLY UNDERSTOOD. MANY FACTORS CAN CONTRIBUTE TO THE ONSET AND PROPAGATION OF CALCIFICATION INCLUDING PATIENT RELATED (E.G. PATIENT AGE, DISEASE STATE, IMMUNE STATUS, AND OTHER CO-MORBIDITIES), PHARMACOLOGICAL, AND INTRINSIC PROPERTIES OF THE VALVE ITSELF. IT IS WIDELY UNDERSTOOD THAT PATIENTS WITH CHRONIC RENAL DISEASE AND PRIOR HISTORY OF CALCIFIC STENOSIS OF THE NATIVE VALVE MAY BE PREDISPOSED TO BIOPROSTHETIC CALCIFICATION. NO FURTHER ACTIONS ARE POSSIBLE WITH THE AVAILABLE INFORMATION.

Description of Event or Problem · 1

IN THIS CASE, IT WAS REPORTED THAT THE PATIENT HAD RE-OPERATION FOR A VALVE-IN-VALVE PROCEDURE AFTER AN IMPLANT DURATION OF 10 YEARS AND 6 MONTHS. THROUGH FOLLOW-UP WITH THE HEALTH-CARE PROVIDER, IT WAS LEARNED THAT THIS PATIENT HAD DEVELOPED PROSTHETIC AORTIC VALVE STENOSIS DUE TO CALCIFICATION. IT WAS DECIDED TO REPLACE THE PATIENT'S AORTIC VALVE VIA TRANSAPICAL TRANSCATHETER VALVE-IN-VALVE INSERTION. THE EDWARDS TRANSCATHETER VALVE WAS IMPLANTED WITHOUT DIFFICULTY. THE PATIENT WAS REPORTED TO HAVE TOLERATED THE PROCEDURE WELL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS REPLACEMENT HEART VALVE DYE EDWARDS LIFESCIENCES 2800

Patients

Seq Age Sex Outcome Treatment
1 77 YR Hospitalization| R