LEAD MODEL 302
Report
- Report Number
- 1644487-2012-02986
- Event Type
- Malfunction
- Date Received
- November 13, 2012
- Date of Event
- October 15, 2012
- Report Date
- October 15, 2012
- Manufacturer
- CYBERONICS, INC.
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE FAILURE OCCURRED BUT DID NOT CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR DEATH.
DEVICE FAILURE IS SUSPECTED BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH.
ON (B)(6) 2012, IT WAS REPORTED THAT THIS VNS PATIENT UNDERWENT GENERATOR REVISION. THE GENERATOR WAS EXPLANTED AND A NEW GENERATOR WAS IMPLANTED; HOWEVER, HIGH IMPEDANCE WAS SEEN. GENERATOR DIAGNOSTICS ON BOTH THE NEW AND OLD GENERATOR WITH THE TEST RESISTOR WERE PERFORMED WITH NORMAL RESULTS. ATTEMPTS TO RE-INSERT THE PIN WERE THREE TIMES WITH THE NEW GENERATOR. DIAGNOSTICS WERE NOT RUN ON THE OLD GENERATOR CONNECTED TO THE LEAD. X-RAYS WERE TAKEN IN THE OPERATING ROOM AND IT WAS VERIFIED THAT SOMETHING WAS WRONG WITH THE LEAD; HOWEVER, THE X-RAYS IMAGES WERE NOT SAVED AND WOULD NOT BE PROVIDED FOR REVIEW. THE NEW GENERATOR WAS PLACED, THE DEVICE WAS LEFT PROGRAMMED OFF, AND THE LEAD WAS NOT REPLACED AT THE TIME OF SURGERY. SURGERY IS LIKELY BUT HAS NOT TAKEN PLACE.
PRODUCT ANALYSIS WAS APPROVED ON (B)(6) 2013. A BREAK WAS IDENTIFIED IN THE POSITIVE COIL. SCANNING ELECTRON MICROSCOPY IMAGES OF THE POSITIVE COIL SHOW THAT PITTING OR ELECTRO-ETCHING CONDITIONS HAVE OCCURRED AT THE BREAK LOCATION. IN ADDITION ABRASIONS IN THE INNER AND OUTER TUBING WERE FOUND. NOTE THAT SINCE THE ELECTRODE ARRAY PORTION WAS NOT RETURNED FOR ANALYSIS, AN EVALUATION AND RESULTING COMMENTARY CANNOT BE MADE ON THAT PORTION OF THE LEAD. OTHER THAN THE ABOVE MENTIONED OBSERVATIONS AND TYPICAL WEAR AND EXPLANT RELATED OBSERVATIONS, NO OTHER ANOMALIES WERE IDENTIFIED IN THE RETURNED LEAD PORTION.
ON (B)(6) 2012, THIS VNS PATIENT UNDERWENT LEAD REVISION. PRE-IMPLANT DIAGNOSTICS INDICATED HIGH IMPEDANCE. THE PATIENT'S GENERATOR WAS LEFT AT THE SETTINGS HE CAME IN TO THE OPERATING ROOM WITH. (THESE SETTINGS ARE INDICATIVE OF A FAULTED SYSTEM DIAGNOSTIC TEST. THIS EVENT IS CAPTURED IN MFR REPORT # 1644487-2013-00002.) POST-IMPLANT DIAGNOSTICS SHOWED PROPER DEVICE FUNCTION. ACCORDING TO THE DOCTORS, THE TOP TWO ELECTRODES OF THE OLD LEAD WERE OFF THE NERVE. THE LEAD WAS RECEIVED ON (B)(4) 2012 AND IS CURRENTLY UNDERGOING PRODUCT ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LEAD MODEL 302 | LEAD | LYJ | CYBERONICS, INC. | 302-20 | 200631 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR |