FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2830622 · Received November 13, 2012

Report

Report Number
2531779-2012-13550
Event Type
Malfunction
Date Received
November 13, 2012
Report Date
October 18, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4):A REVIEW OF THE BLACK BOX SHOWS EVIDENCE OF REBOOTING. VISUAL INSPECTION REVEALED NO DAMAGE TO THE BATTERY CAP OR BATTERY COMPARTMENT. THE BATTERY CAP WAS ABLE TO FULLY TIGHTEN TO THE PUMP WITH NO YELLOW O-RING SHOWING.

Additional Manufacturer Narrative · 1

ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THIS REPORT IS MADE UNDER THE REQUIREMENTS OF THE MEDICAL DEVICE REPORTING REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION ON THE PART OF ANIMAS OF ANY DEFICIENCY IN THE PERFORMANCE OF THE DEVICE.ANIMAS HAS RECEIVED THE PUMP INVOLVED WITH THE COMPLAINT BUT HAVE NOT COMPLETED INVESTIGATIONS. A SUPPLEMENTAL REPORT WILL BE SUBMITTED ONCE INVESTIGATION HAS BEEN COMPLETED.

Description of Event or Problem · 1

ON (B)(6) 2012, THE REPORTER CONTACTED ANIMAS TO REPORT AN INTERMITTENT POWER ISSUE WITH THE PUMP. THE REPORTER CLAIMED THERE WAS NO VISIBLE DAMAGE TO THE PUMP. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT. THERE IS NO INDICATION THAT THE PUMP CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. BASED ON THE PROVIDED INFORMATION, THIS COMPLAINT IS BEING REPORTED BECAUSE THE INTERMITTENT POWER ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. A REPLACEMENT PUMP WAS SENT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 67 YR