HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2012-04998
- Event Type
- Malfunction
- Date Received
- November 13, 2012
- Date of Event
- October 21, 2012
- Report Date
- October 21, 2012
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PATIENT
Narratives
(B)(4). THE PROBLEM WAS CONFIRMED. THE ROOT CAUSE WAS SUPPLY FELL AND DISCONNECTED. THE LABEL REVIEW FOUND THE LABELING ADEQUATE FOR THE USE ERROR IN THIS COMPLAINT.
THE CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REGARDING A CHECK LINES AND BAGS (CLB) ALARM, WHICH OCCURRED ON THE HOMECHOICE (HC) DURING USE, DURING DWELL. THE CAREGIVER (CG) STATED THE HOME PATIENT (HP) HAD FOUR BAGS CONNECTED AND ONLY HAD FLUID ON THE FINAL LINE. THE BAXTER TECHNICAL SERVICE REPRESENTATIVE (TSR) DETERMINED THE HP WAS IN FILL 2 OF 5. THE TSR REVIEWED THE PROGRAMMING AND FOUND IT TO BE GOOD. THE CG STATED A BAG BECAME DISCONNECTED EARLIER. THE TSR INFORMED THE CG THAT WAS THE REASON FOR RUNNING SHORT OF SOLUTION. THE TSR ASSISTED THE CG TO END THERAPY AND DISPOSE OF ALL CURRENT SUPPLIES AND START OVER USING ALL NEW SUPPLIES. THE TSR ALSO ADVISED THE CG NOT TO RECONNECT A BAG DURING THERAPY. THE HC WAS OPERATIONAL. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT. PRODUCT SURVEILLANCE CONTACTED THE HOME PATIENT'S CAREGIVER AND SHE SAID THAT DURING THERAPY THE PATIENT'S BAG ROLLED OFF A TABLE AND ONTO THE GROUND. THAT WAS WHEN THE BAG BECAME DISCONNECTED. SHE HAD ALREADY DISCUSSED WITH THE TSR THE IMPORTANCE OF NOT RECONNECTING A BAG AS IT IS A BREACH IN ASEPTIC TECHNIQUE. SINCE THEN THE PATIENT HAS BEEN COMPLETING THERAPY SUCCESSFULLY. THERE WAS PATIENT INVOLVEMENT BUT NO REPORTED INJURY OR MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | HOME CHOICE |