FDA Adverse Event Malfunction Summary report: N

LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR

MDR report key: 2830426 · Received November 13, 2012

Report

Report Number
3015876-2012-00840
Event Type
Malfunction
Date Received
November 13, 2012
Date of Event
October 19, 2012
Report Date
October 19, 2012
Manufacturer
PHYSIO-CONTROL, INC
Product Code
LDD
PMA / PMN Number
K063119
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND OBSERVED THAT IT TOOK LONGER TO BOOT UP. HOWEVER, THE REPORTED DEVICE RESTARTING ISSUE WAS NOT VERIFIED OR DUPLICATED. PHYSIO COMPLETED OTHER UNRELATED REPAIRS AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. A CONCLUSIVE CAUSE OF THE REPORTED EVENT WAS NOT DETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE KEPT RESTARTING AND FAILED TO BOOT UP PROPERLY. THE DEVICE WOULD NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY IN THE REPORTED CONDITION. THERE WAS NO PATIENT USE ASSOCIATED WITH THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES) LDD PHYSIO-CONTROL, INC 20

Patients

Seq Age Sex Outcome Treatment
1