FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2830417 · Received November 13, 2012

Report

Report Number
2531779-2012-13536
Event Type
Malfunction
Date Received
November 13, 2012
Report Date
November 5, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): THE KEYPAD WAS FOUND TO BE INTACT. THE CONTRAST BUTTON WAS FOUND TO BE UNRESPONSIVE TO PRESSES. THE CONTRAST BUTTON HAS NO EFFECT ON THE PUMP'S INSULIN DELIVERY FUNCTIONS. THE KEYPAD WAS REMOVED AND CONTAMINATION WAS OBSERVED UNDER ALL OF THE BUTTON CONTACTS. UNRELATED TO THE COMPLAINT, THE DISPLAY SCREEN WAS FOUND TO BE FADED.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

THE PATIENT CONTACTED ANIMAS ON (B)(6) 2012 STATING THAT THE UP ARROW AND DOWN ARROW KEYPAD BUTTONS REQUIRED MULTIPLE PRESSES TO ELICIT A RESPONSE. THE PATIENT DENIED ANY DAMAGE TO THE KEYPAD OR ANY EVIDENCE OF MOISTURE IN THE PUMP. THE PATIENT REPORTEDLY WORE THE PUMP IN A POCKET AND DID NOT CLEAN THE PUMP. THERE IS NO ADVERSE EVENT WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED DUE TO THE ALLEGED KEYPAD ISSUES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 34 YR