FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 2830381 · Received November 13, 2012

Report

Report Number
1416980-2012-04981
Event Type
Malfunction
Date Received
November 13, 2012
Date of Event
July 30, 2012
Report Date
October 30, 2012
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION: THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE EVENT DETAILS LEADING TO THE IIPV EVENT WERE CUT OFF. THE CAUSE FOR THE IIPV EVENT WAS UNDETERMINED AS THERE WAS A LACK OF EVIDENCE TO MAKE A DETERMINATION. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.

Description of Event or Problem · 1

A HIGH DRAIN ERROR 106 (NIGHT DRAIN CYCLE SIX) ALARM WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE DEVICE. THE ALARM OCCURRED ON (B)(6) 2012 AT 05:22:13. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1