FDA Adverse Event Malfunction Summary report: N

DIMENSION® CLINICAL CHEMISTRY SYSTEM

MDR report key: 2830375 · Received November 13, 2012

Report

Report Number
2517506-2012-00336
Event Type
Malfunction
Date Received
November 13, 2012
Date of Event
October 14, 2012
Report Date
October 14, 2012
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Product Code
MMI
PMA / PMN Number
K010313
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT IS SAMPLE INTEGRITY. THE ACCOUNT DOES NOT MEET THE TUBE MANUFACTURER'S RECOMMENDATIONS FOR SPIN SPEED AND DURATION. THE IFU FOR DIMENSION CTNI FLEX REAGENT CARTRIDGE CONTAINS THE FOLLOWING INFORMATION: "FOLLOW THE INSTRUCTIONS WITH YOUR SPECIMEN COLLECTION DEVICE FOR USE AND PROCESSING." AND "SPECIMENS SHOULD BE FREE OF PARTICULATE MATTER." THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

A FALSELY ELEVATED TROPONIN I RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN. THE SAMPLE WAS RERUN AND A LOWER RESULT WAS OBTAINED. A CORRECTED REPORTED WAS ISSUED. IT IS UNKNOWN IF PATIENT TREATMENT WAS ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DIMENSION® CLINICAL CHEMISTRY SYSTEM CARDIAC TROPONIN I FLEX® REAGENT CARTRIDGE MMI SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW DA3213

Patients

Seq Age Sex Outcome Treatment
1