FDA Adverse Event Malfunction Summary report: N

PROGRASP (TM) FORCEPS

MDR report key: 2830365 · Received November 13, 2012

Report

Report Number
2955842-2012-00905
Event Type
Malfunction
Date Received
November 13, 2012
Date of Event
October 10, 2012
Report Date
October 16, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE PITCH CABLE WAS FOUND TO BE BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP WAS MISSING FROM CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. NO OTHER DAMAGE FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT. INTUITIVE SURGICAL CONTACTED THE INITIAL REPORTER OF THIS COMPLAINT TO OBTAIN ADDITIONAL INFORMATION CONCERNING THE REPORTED EVENT; HOWEVER, NO ADDITIONAL INFORMATION HAS BEEN PROVIDED AS OF THE DATE OF THIS REPORT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DAVINCI SI PROSTATECTOMY PROCEDURE, THE CUSTOMER NOTED THAT THE WIRES WERE EXPOSED ON THE PROGRASP FORCEPS INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRASP (TM) FORCEPS ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420093-08 M10120718 183

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES