ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2012-13523
- Event Type
- Malfunction
- Date Received
- November 13, 2012
- Report Date
- October 17, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP WAS RETURNED FOR INVESTIGATION AND DEVICE EVALUATION WAS COMPLETED (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THE "CONTRAST" BUTTON IS UNRESPONSIVE AND THE "UP, DOWN, AND OK" BUTTONS ARE INTERMITTENT WHEN PRESSED. THE KEYPAD WAS INTACT AND WHEN REMOVED FOR INVESTIGATION, CONTAMINATION WAS FOUND UNDER ALL BUTTON CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED FOR PRODUCT ANALYSIS INVESTIGATION AND THE EVALUATION REVEALED NONRESPONSIVE BUTTONS AND CONTAMINATION UNDER THE BUTTON CONTACTS. THIS REPORT IS MADE BASED ON THE RESULTS OF AN INVESTIGATION COMPLETED ON (B)(4) 2012.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 25 YR |